Shopping Cart

No products in the cart.

ASTM-E2593 2011

$44.96

E2593-11 Standard Guide for Accuracy Verification of Industrial Platinum Resistance Thermometers

Published By Publication Date Number of Pages
ASTM 2011 14
Guaranteed Safe Checkout
Categories: ,

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

ASTM E2593-11

Historical Standard: Standard Guide for Accuracy Verification of Industrial Platinum Resistance Thermometers

ASTM E2593

Scope

1.1 This guide describes the techniques and apparatus required for the accuracy verification of industrial platinum resistance thermometers constructed in accordance with Specification E 1137/E 1137M and the evaluation of calibration uncertainties. The procedures described apply over the range of 200 °C to 650 °C.

1.2 This guide is not intended to describe the procedures necessary for the calibration of platinum resistance thermometers intended to be used as calibration standards or Standard Platinum Resistance Thermometers. Consequently, calibration of these types of instruments is outside the scope of this guide.

1.3 Industrial platinum resistance thermometers are available in many styles and configurations. This guide does not purport to determine the suitability of any particular design, style, or configuration for calibration over a desired temperature range.

1.4 The evaluation of uncertainties is based upon current international practices as described in ISO/TAG 4/WG 3 Guide to the Evaluation of Uncertainty in Measurement and ANSI/NCSL Z540-2-1997 U.S. Guide to the Expression of Uncertainty in Measurement.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Keywords

accuracy verification; calibration; industrial platinum resistance thermometer; platinum resistance thermometer; standard platinum resistance thermometer; uncertainty; Industrial thermometers; Platinum resistance thermometers (PRT)

ICS Code

ICS Number Code 17.200.20 (Temperature-measuring instruments)

DOI: 10.1520/E2593-11

PDF Catalog

PDF Pages PDF Title
1 Scope
Referenced Documents
Terminology
2 Summary of Guide
Significance and Use
Sources of Error
Apparatus
4 Preparation of UUT
Procedure
TABLE 1
5 Calculation
TABLE 2
6 Report
TABLE 3
TABLE 4
7 Uncertainty
8 TABLE 5
9 Keywords
X1. THERMOMETER WIRE CONFIGURATIONS AND MEASUREMENT SCHEMES
X1.1
X1.2
FIG. X1.1
FIG. X1.2
FIG. X1.3
10 FIG. X1.4
FIG. X1.5
FIG. X1.6
11 X2. RESISTANCE MEASUREMENT INSTRUMENTS AND SCHEMES
X2.1 AC Bridge
X2.2 DC Bridge
X2.3 Digital Thermometer Readout
FIG. X1.7
12 X2.4 Digital Multimeter (DMM)
FIG. X2.1
13 FIG. X2.2
FIG. X2.3
14 FIG. X2.4
FIG. X2.5
ASTM-E2593 2011
$44.96