ASTM-E2593 2023
$37.38
E2593-17(2023) Standard Guide for Accuracy Verification of Industrial Platinum Resistance Thermometers
Published By | Publication Date | Number of Pages |
ASTM | 2023 |
ASTM E2593-17-Reapproved2023
Active Standard: Standard Guide for Accuracy Verification of Industrial Platinum Resistance Thermometers
ASTM E2593
Scope
1.1 This guide describes the techniques and apparatus required for the accuracy verification of industrial platinum resistance thermometers constructed in accordance with Specification E1137/E1137M and the evaluation of calibration uncertainties. The procedures described apply over the range of -200 Ā°C to 650 Ā°C.
1.2 This guide does not intend to describe procedures necessary for the calibration of platinum resistance thermometers used as calibration standards or Standard Platinum Resistance Thermometers. Consequently, calibration of these types of instruments is outside the scope of this guide.
1.3 Industrial platinum resistance thermometers are available in many styles and configurations. This guide does not purport to determine the suitability of any particular design, style, or configuration for calibration over a desired temperature range.
1.4 The evaluation of uncertainties is based upon current international practices as described in JCGM 100:2008 āEvaluation of measurement dataāGuide to the expression of uncertainty in measurementā and ANSI/NCSL Z540.2-1997 āU.S. Guide to the Expression of Uncertainty in Measurement.ā
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Keywords
accuracy verification; calibration; industrial platinum resistance thermometer; platinum resistance thermometer; standard platinum resistance thermometer; uncertainty;
ICS Code
ICS Number Code 17.200.20 (Temperature-measuring instruments)
DOI: 10.1520/E2593-17R23