Shopping Cart

No products in the cart.

ASTM-F2502 2005

$44.96

F2502-05 Standard Specification and Test Methods for Bioabsorbable Plates and Screws for Internal Fixation Implants

Published By Publication Date Number of Pages
ASTM 2005 11
Guaranteed Safe Checkout
Categories: ,

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

1.1 This specification and test methods covers a mechanical characterization reference for hydrolytically degradable polymer resin (from this point on referenced as ‘bioabsorbable’) plates and screws for orthopedic internal fixation.

1.2 This specification establishes common terminology to describe the size and other physical characteristics of bioabsorbable implants and performance definitions related to the performance of bioabsorbable devices.

1.3 This specification establishes standard test methods to consistently measure performance-related mechanical characteristics of bioabsorbable devices when tested under defined conditions of pretreatment, temperature, humidity, and testing machine speed.

1.4 This specification may not be appropriate for all bioabsorbable devices. The user is cautioned to consider the appropriateness of the standard in view of the particular bioabsorbable device and its potential application.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

PDF Catalog

PDF Pages PDF Title
1 Scope
Referenced Documents
Terminology
2 Significance and Use
Materials and Manufacture
FIG. 1
3 Conditioning
Care and Handling
Performance Requirements
Driving Instruments
General Requirements and Performance Considerations
Keywords
4 A1. TEST METHOD FOR DETERMINING THE TORSIONAL PROPERTIES OF BIOABSORBABLE BONE SCREWS
A1.1 Scope
A1.2 Apparatus
A1.3 Sampling
A1.4 General Conditions
A1.5 Procedure
FIG. A1.1
5 A1.6 Report
A1.7 Precision and Bias
A2. TEST METHOD FOR DRIVING TORQUE OF BIOABSORBABLE BONE SCREWS
A2.1 Scope
A2.2 Apparatus
A2.3 Sampling
A2.4 Procedure
FIG. A1.2
FIG. A2.1
6 A2.5 Report
A2.6 Precision and Bias
A3. TEST METHOD FOR AXIAL PULLOUT OF BIOABSORBABLE BONE SCREWS
A3.1 Scope
A3.2 Apparatus
A3.3 Sampling
A3.4 Procedure
7 A3.5 Calculation or Interpretation of Results
A3.6 Report
A3.7 Precision and Bias
FIG. A3.1
8 A4. TEST METHOD FOR SINGLE CYCLE BEND TESTING OF BIOABSORBABLE BONE PLATES
A4.1 Scope
A4.2 Terminology
A4.3 Apparatus
FIG. A4.1
9 A4.4 Sampling
A4.5 Procedure
FIG. A4.2
FIG. A4.3
10 A4.6 Report
A4.7 Precision and Bias
X1. RATIONALE
X1.1
X1.2
11 X1.3
X1.4
X1.5
ASTM-F2502 2005