BS EN ISO 80601-2-12:2023 – TC
$280.87
Tracked Changes. Medical electrical equipment – Particular requirements for basic safety and essential performance of critical care ventilators
Published By | Publication Date | Number of Pages |
BSI | 2023 | 414 |
This document applies to the basic safety and essential performance of a critical care ventilator in combination with its accessories, hereafter referred to as ME equipment: ¾ intended for use in an environment that provides specialized care for patients whose conditions can be life-threatening and who can require comprehensive care and constant monitoring in a professional healthcare facility; NOTE 2 For the purposes of this document, such an environment is referred to as a critical care environment. Ventilators for this environment are considered life-sustaining. NOTE 3 For the purposes of this document, such a critical care ventilator can provide ventilation during transport within a professional healthcare facility (i.e. be a transit-operable ventilator). NOTE 4 A critical care ventilator intended for use in transport within a professional healthcare facility is not considered as an emergency medical services environment ventilator. ¾ intended to be operated by a healthcare professional operator; and ¾ intended for those patients who need differing levels of support from artificial ventilation including for ventilator-dependent patients. A critical care ventilator is not considered to use a physiologic closed-loop-control system unless it uses a physiological patient variable to adjust the artificial ventilation therapy settings. This document is also applicable to those accessories intended by their manufacturer to be connected to a ventilator breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator. NOTE 5 If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 7.2.13 and 8.4.1. NOTE 6 Additional information can be found in IEC 60601-1:2005+AMD1:2012+AMD2:2020, 4.2. This document is not applicable to ME equipment or an ME system operating in a ventilator-operational mode solely intended for patients who are not dependent on artificial ventilation. NOTE 7 A critical care ventilator, when operating in such a ventilator-operational mode, is not considered life-sustaining. This document is not applicable to ME equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. This document does not specify the requirements for: NOTE 8 See ISO/TR 21954 for guidance on the selection of the appropriate ventilator for a given patient. ¾ ventilators or accessories intended for anaesthetic applications, which are given in ISO 80601?2?13; ¾ ventilators or accessories intended for the emergency medical services environment, which are given in ISO 80601-2-84; ¾ ventilators or accessories intended for ventilator-dependent patients in the home healthcare environment, which are given in ISO 80601?2-72; ¾ ventilators or accessories intended for home-care ventilatory support devices, which are given in ISO 80601-2-79 and ISO 80601-2-80; ¾ obstructive sleep a
PDF Catalog
PDF Pages | PDF Title |
---|---|
246 | undefined |
279 | 201. 4.3.101 Additional requirements for essential performance |
280 | 201. 4.3.102 System recovery |
281 | 201. 4.4 Additional requirements for expected service life 201. 4.5 Alternative risk control measures or test methods for ME equipment or ME system 201. 4.6 ME equipment or ME system parts that contact the patient |
282 | 201. 4.11.101 Additional requirements for pressurized gas input 201. 4.11.101.1 Overpressure requirement 201. 4.11.101.2 Compatibility requirement for medical gas pipeline systems |
283 | 201. 5.101 Additional requirements for general requirements for testing of ME equipment 201. 5.101.1 Ventilator test conditions 201. 5.101.2 Gas flowrate and leakage specifications |
284 | 201. 5.101.3 Ventilator testing errors 201. 7.1.101 Information to be supplied by the manufacturer |
285 | 201. 7.2.3 Consult accompanying documents 201. 7.2.4.101 Additional requirements for accessories 201. 7.2.18 External gas source 201. 7.2.101 Additional requirements for marking on the outside of ME equipment or ME equipment parts |
286 | 201. 7.4.3 Units of measurement 201. 7.9.2.1 General |
287 | 201. 7.9.2.1.101 Additional general requirements 201. 7.9.2.2.101 Additional requirements for warnings and safety notices |
288 | 201. 7.9.2.8.101 Additional requirements for start-up procedure 201. 7.9.2.9.101 Additional requirements for operating instructions |
289 | 201. 7.9.2.12 Cleaning, disinfection, and sterilization |
290 | 201. 7.9.2.14.101 Additional requirements for accessories, supplementary equipment, used material 201. 7.9.2.16.101 Additional requirements for reference to the technical description 201. 7.9.3.1.101 Additional general requirements |
291 | 201. 7.9.3.101 Additional requirements for the technical description 201. 9.6.2.1.101 Additional requirements for audible acoustic energy |
293 | 201. 9.101 Additional requirements for suction procedures |
295 | 201. 11.1.2.2 Applied parts not intended to supply heat to a patient |
296 | 201. 11.6.5.101 Additional requirements for ingress of water or particulate matter into ME equipment or ME system 201. 11.6.6 Cleaning and disinfection of ME equipment or ME system |
297 | 201. 11.6.7 Sterilization of ME equipment or ME system 201. 11.7 Biocompatibility of ME equipment and ME systems 201. 11.8.101 Additional requirements for interruption of the power supply/supply mains to ME equipment |
299 | 201. 12.1.101 Volume-control inflation-type |
303 | 201. 12.1.102 Pressure-control inflation-type |
306 | 201. 12.1.103 Other inflation-types |
307 | 201. 12.1.104 Inspiratory volume monitoring 201. 12.1.105 Response of the ventilator to an increase in set oxygen (O2) concentration |
309 | 201. 12.4 Protection against hazardous output 201. 12.4.101 Oxygen monitor |
310 | 201. 12.4.102 Measurement of airway pressure |
311 | 201. 12.4.103 Measurement of expired volume and low volume alarm conditions 201. 12.4.103.1 Ventilators intended to provide a tidal volume >50 ml |
312 | 201. 12.4.103.2 Ventilators intended to provide a tidal volume ≤50 ml |
313 | 201. 12.4.104 Expiratory end-tidal CO2 monitoring equipment |
314 | 201. 12.4.105 Maximum limited pressure protection device 201. 12.4.106 High airway pressure alarm condition and protection device |
315 | 201. 12.4.107 PEEP alarm conditions 201. 12.4.108 Obstruction alarm condition |
316 | 201. 12.4.109 Disconnection alarm condition |
317 | 201. 12.4.110 Protection against inadvertent setting of high airway pressure 201. 12.101 Protection against accidental or unintentional adjustments |
318 | 201. 13.2.101 Additional specific single fault conditions 201. 13.2.102 Failure of one gas supply to a ventilator 201. 13.2.103 Independence of ventilation control function and related risk control measures |
319 | 201. 13.2.104 Failure of functional connection to a ventilator control or monitoring means 201. 14.1 General |
320 | 201. 14.102 Cybersecurity capabilities of critical care ventilators 201. 15.3.5.101 Additional requirements for rough handling 201. 15.3.5.101.1 Shock and vibration (robustness) |
321 | 201. 15.3.5.101.2 Shock and vibration for a transit-operable ventilator during operation |
323 | 201. 15.4.1 Construction of connectors 201. 15.101 Mode of operation 201. 15.102 Delivered oxygen concentration 201. 15.103 Accessory self-check |
324 | 201. 16.1.101 Additional general requirements for ME systems 201. 16.2 Accompanying documents of an ME system 201. 101.1 Protection against reverse gas leakage |
325 | 201. 101.2 Connection to an inlet 201. 101.2.1 Low-pressure hose assembly 201. 101.2.2 Filter 201. 101.3 VBS connectors 201. 101.3.1 General |
326 | 201. 101.3.2 Other named ports 201. 101.3.2.1 Patient-connection port 201. 101.3.2.2 Gas output port and gas return port 201. 101.3.2.3 Emergency intake 201. 101.3.2.4 Flow-direction-sensitive components |
327 | 201. 101.3.2.5 Gas pathway connection port 201. 101.3.2.6 Gas exhaust port 201. 101.3.2.7 Temperature sensor port |
328 | 201. 101.3.2.8 Gas intake port 201. 102.1 General 201. 102.2 Labelling 201. 102.3 Breathing sets 201. 102.4 Water vapour management |
329 | 201. 102.4.1 Humidification system 201. 102.4.2 Heat and moisture exchanger (HME) 201. 102.6 Breathing system filters 201. 102.7 Ventilator breathing systems 201. 102.7.1 Leakage from complete VBS 201. 102.7.2 Non-invasive ventilation |
331 | 201. 105.1 General 201. 105.2 Connection to an electronic health record 201. 105.3 Connection to a distributed alarm system or distributed information system about alarm conditions 201. 105.4 Connection for remote control 201. 106.1 Pressure-volume loops |
332 | 201. 106.2 Flow-volume loops 201. 107.1 Expiratory pause |
333 | 201. 107.2 Inspiratory pause |
334 | 202.4.3.1 Compliance criteria 202.5.2.2.1 Requirements applicable to all ME equipment and ME systems 202.8.1.101 Additional general requirements |
335 | 206.101 Primary operating functions |
337 | 206.102 Training 208.6.3.2.2.2.101 Additional requirements for 1 m (operator’s position) visual alarm signals and information signals 208.6.8.3.101 Additional requirements for global indefinite alarm signal inactivation states 208.6.8.4.101 Additional requirements for termination of alarm signal inactivation |
338 | 208.6.12.2 Operator alarm system logging 208.6.12.3 Responsible organization alarm system logging |
340 | 201.C.101 Marking on the outside of ME equipment, ME systems or their parts 201.C.102 Accompanying documents, general |
341 | 201.C.103 Accompanying documents, instructions for use |
344 | 201.C.104 Accompanying documents, technical description |
348 | AA.1 General guidance AA.2 Rationale for particular clauses and subclauses |
388 | BB.1 Background and purpose |
389 | BB.2 Data definition |