BS ISO 21501-4:2018
$167.15
Determination of particle size distribution. Single particle light interaction methods – Light scattering airborne particle counter for clean spaces
Published By | Publication Date | Number of Pages |
BSI | 2018 | 34 |
This document describes a calibration and verification method for a light scattering airborne particle counter (LSAPC), which is used to measure the size distribution and particle number concentration of particles suspended in air. The light scattering method described in this document is based on single particle measurements. The typical size range of particles measured by this method is between 0,1 μm and 10 μm in particle size.
Instruments that conform to this document are used for the classification of air cleanliness in cleanrooms and associated controlled environments in accordance with ISO 14644‑1 and ISO 14644‑2, as well as the measurement of number and size distribution of particles in various environments.
The following parameters are within the scope of this document:
-
size setting error;
-
counting efficiency;
-
size resolution;
-
false count;
-
maximum particle number concentration;
-
sampling flow rate error;
-
sampling time error;
-
response rate;
-
calibration interval;
-
reporting results from test and calibration.
PDF Catalog
PDF Pages | PDF Title |
---|---|
2 | National foreword |
6 | Foreword |
7 | Introduction |
9 | 1 Scope 2 Normative references 3 Terms and definitions |
10 | 4 Principle |
11 | 5 Basic configuration 6 Requirements 6.1 Size setting error |
12 | 6.2 Counting efficiency 6.3 Size resolution 6.4 False count 6.5 Maximum particle number concentration 6.6 Sampling flow rate error 6.7 Sampling time error 6.8 Response rate 6.9 Calibration interval |
13 | 6.10 Reporting of test and calibration results 7 Test and calibration procedures 7.1 Size setting 7.1.1 Evaluation of size setting error |
14 | 7.1.2 Procedure of size setting |
17 | 7.2 Evaluation of counting efficiency |
18 | 7.3 Evaluation of size resolution |
19 | 7.4 Evaluation of false count 7.5 Estimation of coincidence loss at the maximum particle number concentration 7.6 Evaluation of sampling flow rate error |
20 | 7.7 Evaluation of sampling time error 7.8 Evaluation of response rate |
22 | Annex A (informative) Counting efficiency |
24 | Annex B (informative) Size resolution |
25 | Annex C (informative) False count |
26 | Annex D (informative) Response rate |
27 | Annex E (informative) Procedure for evaluating the uncertainties of the results of the performance tests |
33 | Bibliography |